The determination of ammonium in Kjeldahl digests using the gas-sensing ammonia electrode. Comparison of the direct method with the known-addition method.

作者信息

Nubé M, Van den Aarsen C P, Giliams J P, Hekkens W T

出版信息

Clin Chim Acta. 1980 Jan 31;100(3):239-44. doi: 10.1016/0009-8981(80)90272-7.

Abstract

The efficacy of the ammonia electrode for analysis of the nitrogen content of a large series of Kjeldahl digests was investigated. By using this electrode, two methods for the measurement of ammonium concentrations were compared, the direct method and the known-addition method. When the direct method was used, a marked shift in the electrode potential occurred within a few hours, causing errors of 9-17% in the results. When the ammonium concentrations were calculated from the difference in electrode potential before and after addition of a known amount of an ammonium standard solution (known-addition method), it was possible to carry out reproducible measurements and the shift in the electrode potential did not influence the results. In two series of identical samples the coefficient of variation was respectively 1.45% and 0.80%.

摘要

研究了氨电极用于分析大量凯氏消化液中氮含量的功效。使用该电极,比较了两种测量铵浓度的方法,即直接法和已知添加法。采用直接法时,电极电位在几小时内发生明显偏移,导致结果出现9% - 17%的误差。当通过添加已知量的铵标准溶液前后电极电位的差值来计算铵浓度(已知添加法)时,可以进行可重复的测量,并且电极电位的偏移不影响结果。在两组相同的样品中,变异系数分别为1.45%和0.80%。