Suppr超能文献

患病新生儿血压的快速测量。

Rapid measurement of blood pressure in the sick neonate.

作者信息

Ogata E S, Herring G R

出版信息

Intensive Care Med. 1980;6(1):7-8. doi: 10.1007/BF01683074.

Abstract

We measured mean arterial blood pressure using a hydrostatic technique in 24 sick neonates who had umbilical artery catheters in place. Value for blood pressure obtained with this technique correlated significantly (r = 0.9844, P less than 0.01) with measurements obtained with standard strain gauge and electronic monitor. The method is simple, quick, and safe. It is particularly useful when the infant's size or clinical condition precludes the use of the sphygmomanometer or Doppler ultrasound device or in a setting, such as during transport, when standard electronic means of blood pressure measurement are unavailable.

摘要

我们采用流体静力技术,对24名已插入脐动脉导管的患病新生儿测量了平均动脉血压。用该技术测得的血压值与用标准应变计和电子监测仪测得的数值显著相关(r = 0.9844,P<0.01)。该方法简单、快速且安全。当婴儿的体型或临床状况不适合使用血压计或多普勒超声设备时,或者在如转运期间这种无法使用标准电子血压测量手段的情况下,该方法特别有用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验