测量欧姆电极与组织的接触电阻以及施加于心肌的电压。

Measurement of the ohmic electrode-tissue contact resistance and the voltage applied to myocardium.

作者信息

Benzing G, Iltis R, Helmsworth J A, Schreiber T, Kaplan S

出版信息

J Biomed Eng. 1980 Jan;2(1):3-8. doi: 10.1016/0141-5425(80)90085-0.

Abstract

The electrical impedance of a pulse generator-electrode-heart system is a complex one. By analysis of an equivalent circuit developed for the above, a model was derived to describe the output voltage applied to an implanted electrode. On the basis of the dynamic behavior of the equivalent circuit when a pacing pulse was applied, a method for measuring true ohmic electrode-tissue contact resistance was developed, and a device was constructed to provide an instant digital readout of its value. The value of that risestance varied from 179 to 521 omega in 13 patients. When a variable resistance R, was connected in series with an implanted electrode, and a threshold voltage for every value of R was determined the actual voltage applied to the myocardium was computed for 16 patients.

摘要

脉冲发生器 - 电极 - 心脏系统的电阻抗是一个复数。通过对为上述系统建立的等效电路进行分析,推导出了一个模型,用于描述施加到植入电极的输出电压。根据施加起搏脉冲时等效电路的动态行为,开发了一种测量真实欧姆电极 - 组织接触电阻的方法,并构建了一种装置以即时数字读出其值。在13名患者中,该电阻值在179至521欧姆之间变化。当一个可变电阻R与植入电极串联连接,并确定每个R值的阈值电压时,计算了16名患者施加到心肌的实际电压。