Suppr超能文献

心脏直视手术中的临时起搏器治疗:人体心肌阈值、组织及界面阻抗的变化

Temporary pacemaker treatment in open heart surgery: variation in myocardial threshold, tissue and interface impedances in man.

作者信息

Ohm O J, Mörkrid L, Skagseth E

出版信息

Pacing Clin Electrophysiol. 1979 Mar;2(2):162-74. doi: 10.1111/j.1540-8159.1979.tb05196.x.

Abstract

Myocardial threshold and impedance of adequately insulated multicore metal electrodes (lengths l1 and l2) were investigated in 28 patients undergoing open heart surgery. Increase in current threshold from the pre-to postoperative period was: 607 +/- 102% (mean +/- SEM) with a constant-current pulse generator and 885 +/- 129% with a constant-voltage pulse generator. Tissue impedance (RT - initial impendance) calculated as voltage/current ratio 90 mus into the pulse changed from 564 +/- 34 omega before surgery to a minimum of 134 +/- 7 omega. Thereafter, there was a gradual increase in RT to 162 +/- 9 omega the day of electrode removal. In 25 of 28 patients the minimum values were reached the third to eighth postoperative day. Electrode/tissue interface impedances--Faraday resistance (RF) and Helmholtz capacity (CH)--were calculated from regression analysis of loaded and unloaded electrograms using the method of least squares. The RF showed a fall from 14.7 +/- 1.4 K omega to 5.2 +/- 0.3 K omega, and the CH (20-40 Hz) rose from 6.0 +/- 0.9 mu F to 15.5 +/- 0.8 muF preoperatively to the day of minimum tissue impedance. There were no further changes until the day of electrode removal. A significant positive correlation was found between CH (p < 0.002), current threshold (p < 0.005) and equivalent electrode length [lequ = l1 X l2/(l1 + l2)]. The electrode signal source impedance calculated from RT, RF and CH was of a magnitude not likely to contribute to demand failures. The low postoperative electrode impendance resulted in excessive load on the constant-voltage generator (condenser discharge type), rendering stimulation of the heart with reasonable current values impossible.

摘要

对28例接受心脏直视手术的患者,研究了充分绝缘的多芯金属电极(长度分别为l1和l2)的心肌阈值和阻抗。使用恒流脉冲发生器时,术前到术后电流阈值的增加为:607±102%(平均值±标准误);使用恒压脉冲发生器时为885±129%。以电压/电流比计算的组织阻抗(RT - 初始阻抗)在脉冲开始90微秒时,从术前的564±34欧姆变化到最低的134±7欧姆。此后,RT逐渐增加,在电极移除当天达到162±9欧姆。28例患者中有25例在术后第3至8天达到最小值。电极/组织界面阻抗——法拉第电阻(RF)和亥姆霍兹电容(CH)——采用最小二乘法通过对加载和卸载心电图的回归分析计算得出。RF从14.7±1.4千欧姆降至5.2±0.3千欧姆,CH(20 - 40赫兹)从术前的6.0±0.9微法升至组织阻抗最低时的15.5±0.8微法。直至电极移除当天均无进一步变化。发现CH(p < 0.002)、电流阈值(p < 0.005)与等效电极长度[lequ = l1×l2/(l1 + l2)]之间存在显著正相关。由RT、RF和CH计算得出的电极信号源阻抗大小不太可能导致需求失败。术后电极阻抗较低导致恒压发生器(电容放电型)负载过大,使得无法以合理的电流值刺激心脏。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验