Giachino A A, Cheng M
J Bone Joint Surg Br. 1980 May;62-B(2):227-9. doi: 10.1302/0301-620X.62B2.7364839.
The intensity of scattered radiation in both a forward and a backward direction from a portable fluoroscope during pinning of the neck of the femur was measured by experiments on a cadaver. The intensity decreased rapidly with increasing distance from the flare of the greater trochanter. Hazard to the surgeon's hands from scattered radiation can be greatly reduced by positioning the C-arm of the fluoroscope so that the x-ray beam is directed laterally.
通过在一具尸体上进行实验,测量了在股骨颈固定过程中便携式荧光透视仪向前和向后散射辐射的强度。强度随着距大转子扩大部分距离的增加而迅速降低。通过调整荧光透视仪的C形臂,使X射线束横向照射,可以大大降低散射辐射对外科医生手部的危害。