Suppr超能文献

散射线减少对X射线影像信息含量及患者辐射剂量的影响:定量论证

Effect of reduced scatter on radiographic information content and patient exposure: a quantitative demonstration.

作者信息

Wagner R F, Barnes G T, Askins B S

出版信息

Med Phys. 1980 Jan-Feb;7(1):13-8. doi: 10.1118/1.594773.

Abstract

If a normal screen-film exposure is made using a conventional grid, a reduced exposure may be used with a superior scatter removal device to record the same image information. This theoretical conclusion is not readily demonstrated because of the limited visibility on the underexposed film. In this study radiographs underexposed by a factor of three with a scanning multiple slit assembly (SMSA) were enhanced autoradiographically and found visually to contain more information than properly exposed radiographs obtained under similar conditions using a conventional Bucky grid. This observation was confirmed by contrast scale and noise measurements. Furthermore, the entrance skin exposure of the radiograph using the SMSA was 45% of that required with the grid technique at the same beam quality.

摘要

如果使用传统滤线栅进行常规的屏-片曝光,那么使用一种更高级的散射去除装置时,可以采用减少的曝光量来记录相同的图像信息。由于在曝光不足的胶片上可见度有限,这一理论结论难以轻易得到证实。在本研究中,使用扫描多重狭缝组件(SMSA)将射线照片曝光不足三倍,通过放射自显影增强后,经肉眼观察发现,与在类似条件下使用传统滤线器格栅获得的正常曝光射线照片相比,前者包含更多信息。这一观察结果通过对比度标度和噪声测量得到了证实。此外,在相同的射线质条件下,使用SMSA的射线照片的入射皮肤曝光量仅为使用滤线栅技术时所需曝光量的45%。