Suppr超能文献

反散射格栅对全场数字化乳腺摄影体模图像的影响。

The effect of the antiscatter grid on full-field digital mammography phantom images.

作者信息

Chakraborty D P

机构信息

Department of Radiology, University of Pennsylvania, Philadelphia 19104, USA.

出版信息

J Digit Imaging. 1999 Feb;12(1):12-22. doi: 10.1007/BF03168622.

Abstract

Computer Analysis of Mammography Phantom Images (CAMPI) is a method for making quantitative measurements of image quality. This article reports on a recent application of this method to a prototype full-field digital mammography (FFDM) machine. Images of a modified ACR phantom were acquired on the General Electric Diagnostic Molybdenum Rhodium (GE-DMR) FFDM machine at a number of x-ray techniques, both with and without the scatter reduction grid. The techniques were chosen so that one had sets of grid and non-grid images with matched doses (200 mrads) and matched gray-scale values (1500). A third set was acquired at constant 26 kVp and varying mAs for both grid conditions. Analyses of the images yielded signal-to-noise-ratio (SNR), contrast and noise corresponding to each target object, and a non-uniformity measure. The results showed that under conditions of equal gray-scale value the grid images were markedly superior, albeit at higher doses than the non-grid images. Under constant dose conditions, the non-grid images were slightly superior in SNR (7%) but markedly less uniform (60%). Overall, the grid images had substantially greater contrast and superior image uniformity. These conclusions applied to the whole kVp range studied for the Mo-Mo target filter combination and 4 cm of breast equivalent material of average composition. These results suggest that use of the non-grid technique in digital mammography with the GE-DMR-FFDM unit, is presently not warranted. With improved uniformity correction procedure, this conclusion would change and one should be able to realize a 14% reduction in patient dose at the same SNR by using a non-grid technique.

摘要

乳腺X线摄影体模图像的计算机分析(CAMPI)是一种用于对图像质量进行定量测量的方法。本文报道了该方法最近在一台原型全场数字化乳腺X线摄影(FFDM)机器上的应用。在通用电气诊断钼铑(GE-DMR)FFDM机器上,采用多种X线技术采集了改良ACR体模的图像,包括使用和不使用散射减少格栅的情况。选择这些技术是为了获得一组剂量匹配(200毫拉德)和灰度值匹配(1500)的有格栅和无格栅图像。第三组图像是在两种格栅条件下,以恒定的26 kVp和变化的mAs采集的。对图像的分析得出了与每个目标物体对应的信噪比(SNR)、对比度和噪声,以及一个不均匀性测量值。结果表明,在灰度值相等的条件下,有格栅图像明显更优,尽管其剂量高于无格栅图像。在恒定剂量条件下,无格栅图像在SNR方面略优(7%),但均匀性明显较差(60%)。总体而言,有格栅图像具有显著更高的对比度和更好的图像均匀性。这些结论适用于针对钼-钼靶滤过组合和4厘米平均组成的乳腺等效材料所研究的整个kVp范围。这些结果表明,目前在使用GE-DMR-FFDM设备进行数字化乳腺X线摄影时,不建议使用无格栅技术。随着均匀性校正程序的改进,这一结论将会改变,并且通过使用无格栅技术,在相同SNR的情况下,应该能够使患者剂量降低14%。

相似文献

本文引用的文献

3
Current status of digital mammography.数字化乳腺摄影的现状
Semin Ultrasound CT MR. 1996 Oct;17(5):424-43. doi: 10.1016/s0887-2171(96)90029-3.
5
A multiparameter optimization of digital mammography.数字乳腺摄影的多参数优化
Phys Med Biol. 1995 Nov;40(11):1841-61. doi: 10.1088/0031-9155/40/11/006.