Fryer T B, Corbin S D, Silverberg G D, Schmidt E V, Ream A K
Biotelem Patient Monit. 1978;5(2):88-112.
In the care of patients with abnormal intracranial pressure, it is important to continuously and accurately measure the intracranial pressure so that treatment can be instituted. Strain gage pressure cells have successfully replaced the open-end manometer for chronic use but have lacked the stability for accurate long-term measurements. The system described uses a capacitive pressure transducer with excellent long-term stability. The transducer, electronics, and transmitter package can be implanted to telemeter the pressure transcutaneously, reducing the danger of infection by the elimination of transcutaneous leads. The pressure modulates the frequency of an oscillator in the 8-10 MHz region and provides an easily telemetered signal. To eliminate the problems of periodic battery replacement, the system is energized inductively by a coil placed externally over the implant device, so that the operating life is virtually unlimited.
在颅内压异常患者的护理中,持续准确地测量颅内压以便进行治疗非常重要。应变计压力传感器已成功取代开放式压力计用于长期测量,但缺乏进行精确长期测量的稳定性。所描述的系统使用具有出色长期稳定性的电容式压力传感器。该传感器、电子设备和发射器组件可植入体内,经皮遥测压力,通过消除经皮导线降低感染风险。压力调制8 - 10兆赫兹区域振荡器的频率,并提供易于遥测的信号。为消除定期更换电池的问题,该系统通过置于植入装置外部的线圈进行感应供电,因此使用寿命几乎不受限制。