Schwarz S
J Clin Chem Clin Biochem. 1980 Apr;18(4):215-20. doi: 10.1515/cclm.1980.18.4.215.
A simple computer program for the data processing and quality control of radioimmunoassays is presented. It is written for low cost programmable desk top calculator (Hewlett Packard 97), which can be afforded by smaller laboratories. The untreated counts from the scintillation spectrometer are entered manually; the printout gives the following results: initial data, logit-log transformed calibration points, parameters of goodness of fit and of the position of the standard curve, control and unknown samples dose estimates (mean value from single dose interpolations and scatter of replicates) together with the automatic calculation of within assay variance and, by use of magnetic cards holding the control parameters of all previous assays, between assay variance.
本文介绍了一种用于放射免疫分析数据处理和质量控制的简单计算机程序。它是为低成本的可编程台式计算器(惠普97)编写的,小型实验室也能负担得起。闪烁光谱仪未经处理的计数需手动输入;打印输出结果如下:初始数据、对数-对数转换后的校准点、拟合优度和标准曲线位置的参数、对照和未知样品剂量估计值(单次剂量内插的平均值和重复测量的离散度),同时自动计算分析内方差,并通过使用保存所有先前分析对照参数的磁卡计算分析间方差。