Ullmann V, Husák V, Dubroka L
Eur J Nucl Med. 1978 Jul 1;3(3):197-202. doi: 10.1007/BF00256644.
Several methods for the measurement of dead time and for evaluation of the dependencies required to correct dynamic studies for dead-time losses are described. Two program variants were written to produce time-activity curves in the selected areas of interest of the dynamic studies with computer's correction of dead-time losses; these variants are part of the system of programs for complex processing of scintigraphic studies set up for a Clincom instrument. The first variant performs correction on the basis of the registered count rate in the whole image, while the second variant makes us of the additional reference source positioned on the periphery of the camera visual field.
描述了几种测量死时间以及评估校正动态研究中死时间损失所需相关性的方法。编写了两个程序变体,用于在动态研究的选定感兴趣区域生成时间-活度曲线,并通过计算机校正死时间损失;这些变体是为Clincom仪器建立的闪烁扫描研究复杂处理程序系统的一部分。第一个变体基于整个图像中记录的计数率进行校正,而第二个变体利用位于相机视野周边的附加参考源。