Hudson F R, Davis J B, Whittingham A G
Eur J Nucl Med. 1978 Jul 1;3(3):203-5. doi: 10.1007/BF00256645.
The problem of choosing which collimator to use for imaging a new isotope has been approached by collecting resolution and sensitivity data for a selected group of isotopes. These have been chosen to be readily available and to have generally a single gamma-ray only. Resolution and sensitivity plots for a low energy collimator and a high energy collimator are presented and their use with several isotopes of interest is discussed. The interpretation of recommedations in the literature on the choice of collimators for newly introduced isotopes would be considerably simpler if data in this format were commonly available.
通过收集一组选定同位素的分辨率和灵敏度数据,来解决选择哪种准直器对新同位素进行成像的问题。这些同位素已被选定为易于获取且通常仅具有单一伽马射线。给出了低能准直器和高能准直器的分辨率和灵敏度曲线,并讨论了它们与几种感兴趣的同位素的使用情况。如果以这种格式的数据普遍可用,那么文献中关于为新引入的同位素选择准直器的建议的解释将会简单得多。