用于心肌成像的准直。二。

Collimation for imaging the myocardium. II.

作者信息

Graham L S, Poe N D, Robinson G D

出版信息

J Nucl Med. 1976 Aug;17(8):719-23.

PMID:932817
Abstract

Line-source response functions and modulation transfer functions (MTF) were used to compare the spatial resolutions obtained with an Anger camera system and four different nuclides used as myocardial-imaging agents: 99mTc, 123I, 201Tl, and 43K. The measurements were made with a low-energy converging collimator (LECY, a medium-energy converging collimator (MEC), and a pinhole collimator. The MTF values for 99mTc were very similar for all three collimator types, although the LEC collimator gave slightly higher values at hgih spatial frequencies and had 40% greater sensitivity. Iodine-123 was satisfactorily imaged only with the MEC and pinhole collimators, which in turn yielded MTF values comparable to those measured for 99mTc. Thallium-201 produced MTF curves that were similar for the MEC and pinhole collimators; the curve for th LEC collimator was slightly poorer. All three MTF curves for 201 Tl were inferior to those of 99mTc. For imaging with 43K, only the pinhole collimator provided marginally acceptable spatial resolution.

摘要

使用线源响应函数和调制传递函数(MTF)来比较用安格相机系统以及四种用作心肌显像剂的不同核素(99mTc、123I、201Tl和43K)所获得的空间分辨率。测量是使用低能汇聚准直器(LECY)、中能汇聚准直器(MEC)和针孔准直器进行的。对于所有三种准直器类型,99mTc的MTF值非常相似,尽管低能准直器在高空间频率时给出的值略高且灵敏度高40%。仅使用中能汇聚准直器和针孔准直器时,123I成像效果令人满意,这两种准直器产生的MTF值与99mTc测量值相当。201Tl产生的MTF曲线对于中能汇聚准直器和针孔准直器来说是相似的;低能准直器的曲线稍差。201Tl的所有三条MTF曲线都不如99mTc的曲线。对于用43K成像,只有针孔准直器提供了勉强可接受的空间分辨率。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验