血红蛋白A1色谱分析的温度依赖性以及温度控制分析在糖尿病控制临床评估中的应用。

Temperature dependence of the chromatographic assay of hemoglobin A1 and application of a temperature controlled assay to clinical evaluation of diabetic control.

作者信息

Johnson M W, Dobrea G M, Bendezu R, Wieland R G

出版信息

Clin Chim Acta. 1980 Jul 1;104(3):319-28. doi: 10.1016/0009-8981(80)90389-7.

Abstract

We investigated the temperature dependence of the cation-exchange chromatographic properties of hemoglobin A1 using the "Fast Hemoglobins Test System" obtained from Isolab, Inc., Akron, OH, U.S.A. and a temperature control chamber constructed in our laboratory. Six patient samples, two each from the "normal" range (5--8.5%, "transitional" range (8.5--12.0%) and "uncontrolled diabetic" range (12--20%) described by Isolab (as measured at 24 degrees C) were collected, hemolyzed and assayed according to product insert instructions. Each sample was assayed in triplicate in three separate runs at 20 degrees C, 22 degrees C, 24 degrees C, 26 degrees C and 28 degrees C in the temperature controlled (+/- 0.25 degrees C) chamber. Resultant mean increases in assay results of 0.8, 1.1 and 1.6% hemoglobin A1 per 2 degrees C increase in temperature for the three ranges respectively, indicate that the temperature dependence of the microcolumn assay for hemoglobin A1 is at least partially a function of the relative percentage of glycosylated hemoglobin in the sample. By controlling assay temperature we have reduced interassay variation throughout all concentration ranges by more than one-half to less than 5%. We applied this temperature controlled assay to measurement of hemoglobin A1 levels in 33 non-diabetic subjects and to the screening of hemoglobin A1 levels in 77 diabetic patients regularly attending the diabetic-endocrine outpatient clinic of this hospital. No non-diabetic subject had a HbA1 level above 8.5% and of the diabetic patients, 26% fell within the "normal" range, 43% within the "transitional" range and 31% within the "uncontrolled diabetic" range.

摘要

我们使用从美国俄亥俄州阿克伦市的Isolab公司购得的“快速血红蛋白检测系统”以及我们实验室构建的温度控制箱,研究了血红蛋白A1阳离子交换色谱特性的温度依赖性。采集了6份患者样本,每份样本分别来自Isolab所描述的“正常”范围(5 - 8.5%)、“过渡”范围(8.5 - 12.0%)和“未控制的糖尿病患者”范围(12 - 20%)(均为24℃时的测量值),进行溶血处理,并按照产品说明书进行检测。每份样本在温度控制在(±0.25℃)的箱内,于20℃、22℃、24℃、26℃和28℃下分别进行三次独立检测,每次检测重复三次。结果显示,这三个范围的样本,温度每升高2℃,血红蛋白A1检测结果的平均增加量分别为0.8%、1.1%和1.6%,这表明血红蛋白A1微柱检测的温度依赖性至少部分取决于样本中糖化血红蛋白的相对百分比。通过控制检测温度,我们已将所有浓度范围内的检测间变异减少了一半以上,降至5%以下。我们将这种温度控制检测方法应用于33名非糖尿病受试者血红蛋白A1水平的测量,以及对本院糖尿病 - 内分泌门诊定期就诊的77名糖尿病患者血红蛋白A1水平的筛查。没有非糖尿病受试者的HbA1水平高于8.5%,在糖尿病患者中,26%处于“正常”范围,43%处于“过渡”范围,31%处于“未控制的糖尿病患者”范围。

相似文献

7
Critical factors in the chromatographic measurement of glycohemoglobin (HbA1).
Diabetes. 1980 Jun;29(6):467-74. doi: 10.2337/diab.29.6.467.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验