一种用于测量糖化血红蛋白A1(a+b+c)的快速微量方法。

A rapid micro-scale method for the measurement of Haemoglobin A1(a+b+c).

作者信息

Welch S G, Boucher B J

出版信息

Diabetologia. 1978 Mar;14(3):209-11. doi: 10.1007/BF00429782.

Abstract

A rapid method is described for the measurement of total glycosylated haemoglobins (HbA1(a+b+c). The procedure utilizes 0.05 ml of blood and takes forty minutes to complete manually. Eighty blood samples can be analysed without automation by one person in a day. Each analysis uses less than 2 mg of potassium cyanide, resulting in a method that is both safe and rapid for routine hospital laboratories. The inter-assay coefficient of variation was 4% and that for intra-assay measurements 3%, over the range 5-20%, HbA1(a+b+c). The method confirmed that the level of HbA1(a+b+c) is elevated in imperfectly controlled diabetics. Amongst patients with blood glucose levels of less than 10 mmol/l the mean level of HbA1(a+b+c) was found to be 8.5%; samples from 14 known diabetics gave a mean value of 10.9%, whereas 17 known non-diabetic samples gave a mean value of 8.3%. In the group of samples from 27 diabetic individuals with blood glucose levels above 10 mmol/l the mean level of HbA1(a+b+c) was found to be 13.5%.

摘要

本文描述了一种快速测定总糖化血红蛋白(HbA1(a+b+c))的方法。该方法只需0.05毫升血液,手动操作需40分钟完成。一个人在一天内无需自动化设备即可分析80份血样。每次分析使用的氰化钾少于2毫克,因此该方法对医院常规实验室来说既安全又快速。在HbA1(a+b+c)含量为5%-20%的范围内,批间变异系数为4%,批内测量变异系数为3%。该方法证实,血糖控制不佳的糖尿病患者HbA1(a+b+c)水平升高。在血糖水平低于10毫摩尔/升的患者中,HbA1(a+b+c)的平均水平为8.5%;14份已知糖尿病患者的样本平均值为10.9%,而17份已知非糖尿病患者的样本平均值为8.3%。在27名血糖水平高于10毫摩尔/升的糖尿病患者的样本组中,HbA1(a+b+c)的平均水平为13.5%。