Kiprov D, Stoilova I
Eksp Med Morfol. 1980;19(1):55-63.
A modification of an apparatus for undirect bloodless measurement of the arterial blood pressure (BP) in the tail of small laboratory animals (rats) is described. The device consists of a 3-channel ECG apparatus (C3a), connected with a blood pressure recorder(R67z/32) with an osciliscope (R17e) for visual checking of blood pressure. Criteria as well as possibility for measuring not only the systolic, but also the diastolic level of the blood pressure are indicated due to the great sensitivity of the device. It is pointed out that the coefficient of variation (3,58% for the systole and 5,95% for the diastole) is very low which confirms the reliability of the method. The results of a continuous (12 months) dynamic follow-up of the systolic and diastolic blood pressure of normotensive (NWR) and spontaneously hypertensive rats (SHR) are also reported. The authors conclude that the results could be used as normatyves and as indices for the range of the systolic and diastolic blood pressure of NWR and SHR aged 1 to 12 months, since no data about the diastolic blood pressure of rats are available so far in the literature.
描述了一种用于对小型实验动物(大鼠)尾部动脉血压(BP)进行无创无血测量的装置的改进。该装置由一台三通道心电图仪(C3a)组成,与一台带有示波器(R17e)的血压记录仪(R67z/32)相连,用于直观检查血压。由于该装置灵敏度高,不仅指出了测量收缩压的标准和可能性,还指出了测量舒张压的标准和可能性。指出变异系数(收缩期为3.58%,舒张期为5.95%)非常低,这证实了该方法的可靠性。还报告了对正常血压大鼠(NWR)和自发性高血压大鼠(SHR)的收缩压和舒张压进行连续(12个月)动态随访的结果。作者得出结论,这些结果可作为1至12个月龄NWR和SHR收缩压和舒张压范围的标准和指标,因为目前文献中尚无关于大鼠舒张压的数据。