Cosyns-Duyck M C, Cruyl A A, De Leenheer A P, De Schryver A, Huys J V, Belpaire F M
Biomed Mass Spectrom. 1980 Feb;7(2):61-4. doi: 10.1002/bms.1200070205.
Fluorouracil in human plasma of a concentration as low as 2 ng ml-1 can be assayed with the use of gas-liquid chromatography and a multiple ion detector. Fluorouracil is isolated from plasma by anion exchange extraction. The extract is butylated and the derivative obtained extracted into a mixture of cyclohexane + dichloromethane (95:5). An aliquot is submitted to combined gas-liquid chromatography mass spectrometry. The use of chlorouracil or oxygen-18 labelled fluorouracil as internal standard is discussed. Linearity is proven from 100 microgram to 5 ng of fluorouracil per ml plasma. The coefficients of variation are 7.2 and 8.4% for within-run and between-run precision, respectively, at the 50 ng ml-1 level. The biological applicability of our procedure is demonstrated by analysing plasma samples obtained at different time intervals from three patients given 1 g of fluorouracil once intravenously and once orally.
使用气液色谱法和多离子检测器可测定人血浆中低至2 ng/ml浓度的氟尿嘧啶。通过阴离子交换萃取从血浆中分离出氟尿嘧啶。提取物进行丁基化处理,所得衍生物萃取到环己烷+二氯甲烷(95:5)的混合物中。取一份等分试样进行气液色谱 - 质谱联用分析。讨论了使用氯尿嘧啶或氧 - 18标记的氟尿嘧啶作为内标。已证明在每毫升血浆中氟尿嘧啶含量从100微克至5纳克范围内呈线性。在50 ng/ml水平下,批内精密度和批间精密度的变异系数分别为7.2%和8.4%。通过分析从三名患者身上采集的血浆样本(这些患者分别静脉注射和口服1克氟尿嘧啶,且在不同时间间隔采集样本),证明了我们方法的生物学适用性。