Benabid A L, Balme L, Persat J C, Belleville M, Chirossel J P, Buyle-Bodin M, de Rougemont J, Poupot C
TIT J Life Sci. 1978;8(1-2):59-68.
In previous studies, the authors have shown the importance of impedance recording by stereotactic procedures for brain tumor detection. In this study they report the principles and the first results of an attempt to obtain the impedance value of any given point inside the brain by an external way. The experimental device is mainly made of a set of 128 electrodes, surrounded by an electrode guarding monitored by an operational amplifier the purpose of which is to made the electric field uniform and to enable a focused measure of the impedance by each of the 128 electrodes. The effect of this electrode guarding is demonstrated, and the present sensitivity of the prototype is shown on various obstacles. conductive or non-conductive; the influence of an impedance barrier on this sensitivity is the project of an impedometric scanner is described, and the problems due to the non-linear propagation of electric currents and to the anisotropic properties of the brain tissue are evaluated.
在先前的研究中,作者已经表明立体定向程序进行阻抗记录对脑肿瘤检测的重要性。在本研究中,他们报告了通过外部方式获取脑内任意给定点阻抗值的尝试的原理和初步结果。实验装置主要由一组128个电极组成,周围有一个由运算放大器监控的电极保护装置,其目的是使电场均匀,并使128个电极中的每一个都能对阻抗进行聚焦测量。展示了这种电极保护的效果,并在各种障碍物(导电或不导电)上展示了原型的当前灵敏度;评估了阻抗屏障对这种灵敏度的影响,描述了阻抗扫描仪的方案,以及评估了由于电流的非线性传播和脑组织的各向异性特性所导致的问题。