Magilligan D J, Lewis J W, Jara F M, Lee M W, Alam M, Riddle J M, Stein P D
Ann Thorac Surg. 1980 Sep;30(3):259-66. doi: 10.1016/s0003-4975(10)61254-4.
From October, 1971, to October, 1979, 490 patients with 560 porcine bioprosthetic valves were discharged from the hospital. During these 8 years, 23 valves were removed because of failure due to spontaneous degeneration. Bioprosthetic valve survival without degeneration was at 4 years, 98.9% +/- 86 (standard error); at 5 years, 96.4% +/- 1.3; at 6 years, 90.8% +/- 2.4; and at 7 years, 84.2% +/- 3.7. There was no difference in degeneration observed with regard to sex, valve position, or whether the valves were rinsed with antibiotics prior to implantation. There was an increase in degeneration in patients 35 years old and younger compared with those more than 35 years old (p = 0.0001). Valve failure was gradual, and valve changes were noted by echocardiogram and phonocardiogram prior to actual failure. Specific factors leading to degeneration require further investigation.
从1971年10月至1979年10月,490例患者植入了560枚猪生物瓣后出院。在这8年期间,有23枚瓣膜因自然退变失败而被取出。无退变的生物瓣生存率在4年时为98.9%±1.86(标准误);5年时为96.4%±1.3;6年时为90.8%±2.4;7年时为84.2%±3.7。在性别、瓣膜位置或植入前瓣膜是否用抗生素冲洗方面,未观察到退变差异。与35岁以上患者相比,35岁及以下患者的退变率有所增加(p = 0.0001)。瓣膜失败是渐进性的,在实际失败前通过超声心动图和心音图可观察到瓣膜变化。导致退变的具体因素需要进一步研究。