Larkin H, Hunyor S N
J Electrocardiol. 1980 Oct;13(4):347-51. doi: 10.1016/s0022-0736(80)80086-0.
Intra-individual precordial voltage variation was examined in serial 12 lead electrocardiograms (ECGs) performed at 10 minute and 24 hour intervals in sixteen young, healthy males forming two age matched groups. Significant variation was found in repeat ECGs at both periods. When precordial electrodes remained in situ between serial 10 minute recordings variation was reduced by approximately 60% We conclude that significant precordial voltage variation is present in serial electrocardiography, even when performed over the short term. Alteration in precordial electrode placement accounts for the major proportion of variation and this may be sufficiently large to interfere with the accurate interpretation of serial precordial voltage changes in an individual subject.
在16名年轻健康男性中,以10分钟和24小时为间隔进行系列12导联心电图(ECG)检查,以检测个体内心前区电压变化。这16名男性形成两个年龄匹配组。在两个时间段的重复心电图中均发现了显著变化。当在连续的10分钟记录之间心前区电极保持原位时,变化减少了约60%。我们得出结论,即使在短期内进行系列心电图检查,心前区电压也会出现显著变化。心前区电极位置的改变占变化的主要部分,其变化可能足够大,足以干扰对个体受试者系列心前区电压变化的准确解读。