[Five years' experience with an epimyocardial screw-in electrode in cardiac pacemakers].

作者信息

Sege D, Reutter F W, Engel U R, Babotai J

出版信息

Helv Chir Acta. 1980 Jun;47(1-2):237-41.

PMID:7440203
Abstract

197 myocardial sutureless screw-in electrodes were implanted between the years 1973 and 1978. A subxyphoidal transmediastinal route and general anaesthesia were used. All patients had initially good pacing results. In 93% the electrode was placed into the right, in 7% into the left ventricle. The peroperative mortality rate was 0, peroperative complications were seen in 1%. Of the 19% of the complications which needed some kind of a treatment, 7.6% were confined to the lead and 11.4% were other complications. The lead, placed into the right ventricle, had to be replaced in five patients (2.7%) because of exit block or sensing problems. The postpericardiotomy syndrome was the most frequent complication but it did not impair the function of the pacemaker. 4% of the patients died within 30 days after the implantation. In 46 patients the pacemaker was replaced after a mean time of 32,5 months. The measurement of the threshold rose from a mean of 0.8 mAmp to 3.2 mAmp and the mean resistance was 530 Ohms. In 44 patients the electrode could be reused.

摘要

1973年至1978年间植入了197枚心肌无缝拧入式电极。采用剑突下经纵隔途径和全身麻醉。所有患者最初的起搏效果良好。93%的电极置于右心室,7%置于左心室。手术死亡率为0,手术并发症发生率为1%。在需要某种治疗的19%的并发症中,7.6%局限于导线,11.4%为其他并发症。由于出口阻滞或感知问题,5例(2.7%)置于右心室的导线需要更换。心包切开术后综合征是最常见的并发症,但不影响起搏器功能。4%的患者在植入后30天内死亡。46例患者平均在32.5个月后更换了起搏器。阈值测量从平均0.8毫安升至3.2毫安,平均电阻为530欧姆。44例患者的电极可重复使用。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验