超声心动图的可重复性。一项评估系列超声心动图测量变异性的研究。

Reproducibility of echocardiography. A study evaluating the variability of serial echocardiographic measurements.

作者信息

Pietro D A, Voelkel A G, Ray B J, Parisi A F

出版信息

Chest. 1981 Jan;79(1):29-32. doi: 10.1378/chest.79.1.29.

Abstract

In this study, serial reproducibility was determined on standard M-mode (M-mode) and two dimensional guided M-mode (2-D M-mode) records for ten normal volunteers and 20 patients. Overall intraobserver variability in the normal group was 518 +/- 1.8 percent (M-mode) and 3.1 +/- 0.8 percent (2-D M-mode), and in the patient group 3.1 +/- 0.6 percent (M-mode) and 4.7 +/- 0.7 percent (2-D M-mode) (P = NS). Variability on serial examination, however, was twofold to threefold greater. In the normal group, reproducibility varied by 10.8 +/- 1.6 percent (M-mode) and 9.4 +/- 1.2 percent (2-D M-mode), while in the patient group, it was 8.7 +/- 0.9 percent (M-mode) and 9.4 +/- 0.7 percent (2-D M-mode). The lowest serial variability achieved was for the diastolic LV dimension on serial M-mode examination (4 percent); the largest variability pertained to measurement of the interventricular septum (14 percent). Other structures had intermediate variation in reproducibility. This study emphasizes that establishing performance variability for an echocardiographic laboratory is an important part of the interpretation of serial records.

摘要

在本研究中,对10名正常志愿者和20名患者的标准M型(M-mode)和二维引导M型(2-D M-mode)记录进行了系列可重复性测定。正常组观察者内总体变异性在M型为5.18±1.8%,在二维M型为3.1±0.8%;患者组在M型为3.1±0.6%,在二维M型为4.7±0.7%(P=无显著性差异)。然而,系列检查的变异性要大两到三倍。在正常组中,重复性变化在M型为10.8±1.6%,在二维M型为9.4±1.2%;而在患者组中,M型为8.7±0.9%,二维M型为9.4±0.7%。系列M型检查中左心室舒张期内径的系列变异性最低(4%);最大变异性与室间隔测量有关(14%)。其他结构的重复性变化处于中间水平。本研究强调,确定超声心动图实验室的性能变异性是系列记录解读的重要组成部分。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验