犬的多普勒超声心动图:测量变异性与可重复性

Doppler echocardiography in the dog: measurement variability and reproducibility.

作者信息

Dukes-McEwan Joanna, French Anne T, Corcoran Brendan M

机构信息

Hospital for Small Animals, Department of Veterinary Clinical Studies, Royal (Dick) School of Veterinary Studies, The University of Edinburgh, Easter Bush Veterinary Centre, Midlothian, Scotland, UK.

出版信息

Vet Radiol Ultrasound. 2002 Mar-Apr;43(2):144-52. doi: 10.1111/j.1740-8261.2002.tb01662.x.

Abstract

Serial Doppler echocardiographic examinations were carried out in random order on six boxer dogs, on 3 separate days, by two experienced Doppler echocardiographers, to assess measurement variability and reproducibility of 65 parameters. Large numbers of parameters exhibited significant differences for each of the categories of intraobserver, interobserver, interday and interoperator. The coefficients of variation for all parameters measured ranged from 5.03 to 46.43%, but most were less than 20%. In general, least variation was found for the intraobserver category, and the best reproducibility for M-mode and left ventricular volumetric data. The worst reproducibility was found for tricuspid inflow and pulmonary venous flow measurements. The results of this study suggest differences greater than 20% for serial scans must be achieved to document genuine change, although the specific data should be consulted. Furthermore, variability and reproducibility are improved if a single experienced operator/observer acquires and measures serial scans.

摘要

由两位经验丰富的多普勒超声心动图医师在3个不同日期,以随机顺序对6只拳师犬进行系列多普勒超声心动图检查,以评估65项参数测量的变异性和可重复性。大量参数在观察者内、观察者间、日间和操作者间的每一类中均显示出显著差异。所测量的所有参数的变异系数范围为5.03%至46.43%,但大多数小于20%。一般来说,观察者内类别的变异最小,M型和左心室容积数据的可重复性最佳。三尖瓣流入和肺静脉血流测量的可重复性最差。本研究结果表明,尽管应参考具体数据,但要记录真正的变化,系列扫描的差异必须达到20%以上。此外,如果由单一经验丰富的操作者/观察者获取并测量系列扫描,则变异性和可重复性会得到改善。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验