Quantitative and qualitative comparison of two microfocus-tube imaging systems.

作者信息

Gould R G, Genant H K

出版信息

Radiology. 1981 Jan;138(1):195-201. doi: 10.1148/radiology.138.1.7455083.

Abstract

Two microfocus-tube magnification systems were compared: one used an RSI tube (stationary anode, Pierce-gun cathode) and the other used a GE tube (rotating anode, biased cathode). Imaging parameters for body parts of three thicknesses, the hand (low), knee (medium), and hip (high), were considered at a magnification of 2. The focal-spot MTF was measured on and off the central ray of the beam, as well as beam quality and the intensity of scatter in the image plane. Patient exposures were also determined. While both systems gave high-quality images at acceptable exposure levels, the RSI unit was judged superior for body parts up to 15 cm thick. For thicker parts, scatter degraded contrast with the RSI unit, making the GE unit superior.

摘要

对两种微焦点管放大系统进行了比较

一种使用RSI管(固定阳极,皮尔斯枪式阴极),另一种使用GE管(旋转阳极,偏置阴极)。在放大倍数为2的情况下,考虑了三种厚度身体部位(手部(薄)、膝盖(中等)和臀部(厚))的成像参数。在光束中心射线的内外测量焦点MTF,以及光束质量和图像平面中的散射强度。还确定了患者的照射剂量。虽然两种系统都能在可接受的照射水平下提供高质量图像,但对于厚度达15厘米的身体部位,RSI装置被判定更优。对于更厚的部位,散射会降低RSI装置的对比度,使得GE装置更具优势。