Parametrization of head-scatter factors for rectangular photon fields using an equivalent square formalism.

作者信息

Yu M K, Murray B, Sloboda R

机构信息

Cross Cancer Institute, Edmonton, Alberta, Canada.

出版信息

Med Phys. 1995 Aug;22(8):1329-32. doi: 10.1118/1.597617.

Abstract

Head-scatter factors of symmetric square and rectangular fields (field center on the central beam axis) defined by the upper (Y) and lower (X) jaws for 6 and 15 MV photon beams from 2300CD and 600C accelerators (Varian Associates, Inc., Palo Alto, CA) were measured, as well as those for fields shaped by the Y jaws and the multileaf collimator (MLC) of the 2300CD. For rectangular fields, the head-scatter factor for the field (x = a and y = b) was different from that for the field (y = a and x = b). This difference was 2% -3% for fields defined by conventional collimators when [formula, see text] was large, and became 4%-5% when the MLC and YT jaws were used to shape the fields with the X jaws retracted. In order to calculate values for head-scatter factors of rectangular fields accurately using an equivalent square formalism, the side of the equivalent square should be obtained with different weights for lower and upper jaws, as proposed by Vadash and Bjärngard [Med. Phys. 20, 733-734 (1993)]. Our measurements demonstrate that the relative weight (G) of upper and lower jaws is strongly dependent on their distances from the x-ray source, while the beam energy has little effect on the value of G. We further show that G can be calculated simply from these distances. An analytical representation for head-scatter factors of square and rectangular fields is also developed in this paper. The quality of this representation was judged by the root-mean-square (rms) deviation from measured head-scatter factors, which ranged from 0.11%-0.27%.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验