Accuracy of migration analysis in hip arthroplasty. Digitized and conventional radiography, compared to radiostereometry in 51 patients.

作者信息

Malchau H, Kärrholm J, Wang Y X, Herberts P

机构信息

Department of Orthopedics, Sahlgrenska University Hospital, Gothenburg, Sweden.

出版信息

Acta Orthop Scand. 1995 Oct;66(5):418-24. doi: 10.3109/17453679508995578.

Abstract

We assessed the accuracy of migration measurements on conventional and digitized radiographs of total hip arthroplasties by comparing the results with radiostereometry (RSA). 4 stem and 3 acetabular designs were studied. 2 of the stem designs and 1 cup design were uncemented. 180 manual and 202 digitized measurements were done on 120 conventional radiographic examinations. The readings on digitized radiographs did not differ from the manual measurements on the same radiographs. A comparison of the measurements from plain radiographs and with RSA of the femoral stems revealed an accuracy (absolute mean + 2 SD) of between 3.9 and 12.3 mm, depending on the choice of landmarks. The greatest accuracy was obtained by using tantalum markers inserted into the greater or the lesser trochanter and the shoulder of the stem. The most medial point of the lesser trochanter proved to be the best bony landmark. Measurements, including both the center of the femoral head and the greater trochanter, were associated with poor accuracy. The accuracy as regards horizontal cup migration varied from 4.4 to 6.5 mm and the accuracy as regards vertical migration varied between 4.4 and 6.3 mm. The intraobserver error (2 SD) ranged from 1.6 mm to 5.6 mm, The corresponding figures for the inter-observer error were 2.6 mm and 6.6 mm, respectively. One of the cemented cup designs was associated with inferior accuracy. Stem design did not affect the accuracy.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验