Frazier A, Yan D, Du M, Wong J, Vicini F, Matter R, Joyce M, Martinez A
William Beaumont Hospital, Department of Radiation Oncology, Royal Oak Campus, MI 48073, USA.
Int J Radiat Oncol Biol Phys. 1995 Dec 1;33(5):1247-56. doi: 10.1016/0360-3016(95)00273-1.
The purpose of this study is to quantify and compare retrospectively the effects of treatment setup variation on beam's eye view (BEV) dosimetry for radiation therapy using a multileaf collimator (MLC) vs. cerrobend block.
A study was performed on a group of 18 patients with cancer of the head and neck, lung, and pelvis who were treated with irregularly shaped fields. The BEV dosimetry of the fields shaped with cerrobend blocks and the MLC was measured with films at the depth of dose prescription in a solid water phantom. A "one-half-leaf" insertion convention was used to shape the MLC. In addition, an average of 15 sequential daily port films was taken per patient during the course of radiotherapy. The port films were aligned with the prescription film for each patient. Systematic error and random error of treatment setup for each patient were calculated. The effects of setup variation were incorporated by convolving the patient portal imaging data with the corresponding BEV film dosimetry. Two parameters were used to quantify the BEV dosimetry. First, the field penumbra width was calculated, which represented the average of the normal separations between 20 and 80% isodose lines along the prescription outline. Second, the ratio of areas covered by the 90 and 20% isodose lines, A90/20, was determined. The BEV dosimetry was then characterized with and without the effects of treatment setup variation. In addition, the difference in BEV dosimetry between the cerrobend block and the MLC was used to estimate the corresponding changes in tumor control probability (TCP). These changes were also compared to the changes in TCP for the treatment with or without the effects of random setup variation.
With or without daily setup variation, the use of cerrobend block was more favorable than the MLC in terms of the field penumbra width and A90/20 for all treatment sites. In the absence of daily variation, the MLC field penumbra width was on average 1.3 mm larger than that of the cerrobend block, and 0.9 mm larger in the presence of daily setup variation. Similarly, the ratio A90/20 of the cerrobend block was on average 0.03 larger than that of MLC without daily setup variation, and 0.02 with daily setup variation. The difference in field penumbra width and A90/20 between the MLC and the cerrobend block was slightly reduced due to the effects of daily setup variation. For both the cerrobend block and the MLC, daily setup variation produced a significant increase in the field penumbra width, 2.3 mm for the cerrobend block and 1.9 mm for the MLC, and a decrease in the A90/20, 0.06 for the former and 0.05 for the latter. The change due to the daily setup variation was about a factor of 2 larger than the changes due to replacing the cerrobend block with the MLC. Using the TCP model, the change in TCP due to the daily setup variation was more than a factor of 3 larger than the change in TCP due to replacing the cerrobend block with the MLC. It was noted that the average changes in the penumbra, the A90/20 and the TCP calculated for the patient population did not adequately describe the changes for the individual patient.
Our results do not show significant dosimetric differences between the MLC and the cerrobend block in conventional radiation treatment, whether or not daily setup variation was taken into consideration. The effects of daily setup variation alone produced a larger dosimetric change. The same results were obtained when the data were applied to calculate changes in TCP. For optimal radiation therapy, efforts should be concentrated on reducing daily setup variation. Our results also demonstrate the importance of frequent evaluation of MLC treatment using electronic portal imaging devices.
本研究的目的是回顾性地量化和比较使用多叶准直器(MLC)与铈弯曲挡块进行放射治疗时,治疗摆位变化对射野视观(BEV)剂量学的影响。
对一组18例头颈部、肺部和骨盆部癌症患者进行了研究,这些患者接受了不规则野的治疗。在固体水模体中,于剂量处方深度处用胶片测量铈弯曲挡块和MLC形成的射野的BEV剂量学。采用“半叶插入”惯例来塑形MLC。此外,在放射治疗过程中,每位患者平均每天拍摄15张连续的射野片。将射野片与每位患者的处方片对齐。计算每位患者治疗摆位的系统误差和随机误差。通过将患者射野成像数据与相应的BEV胶片剂量学进行卷积,纳入摆位变化的影响。使用两个参数来量化BEV剂量学。首先,计算射野半值层宽度,它代表沿处方轮廓20%和80%等剂量线之间正常间距的平均值。其次,确定90%和20%等剂量线覆盖面积的比值A90/20。然后对有和没有治疗摆位变化影响的BEV剂量学进行表征。此外,铈弯曲挡块和MLC之间BEV剂量学的差异用于估计肿瘤控制概率(TCP)的相应变化。还将这些变化与有或没有随机摆位变化影响的治疗中TCP的变化进行比较。
无论有无每日摆位变化,在所有治疗部位,就射野半值层宽度和A90/20而言,使用铈弯曲挡块比MLC更有利。在没有每日变化时,MLC射野半值层宽度平均比铈弯曲挡块大1.3 mm,在有每日摆位变化时大0.9 mm。同样,铈弯曲挡块的A90/20比值在没有每日摆位变化时平均比MLC大0.03,在有每日摆位变化时大0.02。由于每日摆位变化的影响,MLC和铈弯曲挡块之间射野半值层宽度和A90/20的差异略有减小。对于铈弯曲挡块和MLC两者,每日摆位变化均使射野半值层宽度显著增加,铈弯曲挡块增加2.3 mm,MLC增加1.9 mm,A90/20减小,前者减小0.06,后者减小0.05。每日摆位变化引起的变化比用MLC替代铈弯曲挡块引起的变化大约大2倍。使用TCP模型,每日摆位变化引起的TCP变化比用MLC替代铈弯曲挡块引起的TCP变化大3倍多。值得注意的是,为患者群体计算的半值层、A90/20和TCP的平均变化并不能充分描述个体患者的变化。
我们的结果表明,在传统放射治疗中,无论是否考虑每日摆位变化,MLC和铈弯曲挡块之间在剂量学上没有显著差异。仅每日摆位变化的影响就产生了更大的剂量学变化。将数据应用于计算TCP变化时也得到了相同的结果。为了实现最佳放射治疗,应集中精力减少每日摆位变化。我们的结果还证明了使用电子射野成像设备频繁评估MLC治疗的重要性。