[超声血管造影:慢血流的体模测量]

[Ultrasound angiography: phantom measurements of slow blood flow].

作者信息

Weskott H P, Knuth C

机构信息

Medizinische Klinik II, Städtisches Krankenhaus Siloah, Hannover.

出版信息

Bildgebung. 1995 Sep;62(3):189-92.

PMID:7496115
Abstract

Power Doppler ultrasound is a color-coded Doppler ultrasound method based on the strength of the Doppler shift signal. The sensitivity for slow flow was tested with different flow phantoms. The lowest detectable velocity was 0.04 cm/s in a water- or oil-filled tank at an angle of 45 degrees and 0.24 cm/s in an oil medium at an angle of 86 degrees. The pulse repetition frequency (PRF) was 100 kHz at the lowest detectable velocities, the acquisition time depended mostly on the attenuation of the medium. The new method depended very little on the angle of insonation. Under ideal conditions detectable flow velocities are within the range of mean capillary flow velocity.

摘要

能量多普勒超声是一种基于多普勒频移信号强度的彩色编码多普勒超声方法。使用不同的血流模拟装置测试了对缓慢血流的敏感性。在充满水或油的水箱中,45度角时最低可检测速度为0.04厘米/秒,在油介质中86度角时为0.24厘米/秒。在最低可检测速度下,脉冲重复频率(PRF)为100千赫,采集时间主要取决于介质的衰减。这种新方法对声束入射角度的依赖性很小。在理想条件下,可检测的血流速度在平均毛细血管血流速度范围内。