Suppr超能文献

Comparison of different protocols in plateletpheresis with the haemonetics MCS 3p blood cell separator with respect to parameters of product quality.

作者信息

Moog R, Müller N

机构信息

Institut für Transfusionsmedizin, Universitätsklinikum Essen, Germany.

出版信息

Infusionsther Transfusionsmed. 1995 Aug;22(4):244-8. doi: 10.1159/000223132.

Abstract

BACKGROUND

In thrombocytapheresis, intermittent flow cell separators produce platelet concentrates (PC) with a higher leucocyte contamination than continuous flow blood cell separators. The new discontinuous flow cell separator Haemonetics MCS 3p offers a low-leucocyte solution for PC. The quality of PC obtained by the MCS 3p was investigated in this study.

DESIGN

Prospective study.

SETTING

Haemapheresis Unit of a University Clinic.

PATIENTS

Healthy blood donors from the haemapheresis unit.

MATERIALS AND METHODS

Platelet (PLT) yield, separation efficiency and white blood cell (WBC) contamination were studied in three different protocols. Two protocols used a haemocalculator, which calculated the target volume based on the donor's physical characteristics and the desired PLT yield for the procedure. Protocol I used 3,000 ml as target process volume and protocol II 3.3 x 10(11) as desired PLT yield. Protocol III was used without haemocalculator. Glucose, lactate, lactate dehydrogenase (LDH), morphology score and pH value were analysed to investigate the quality of the PC.

RESULTS

Platelet yield and separation efficiency were not statistically different in the three protocols. Leucocyte contamination was lowest in the protocol without haemocalculator (median: 3.15 x 10(6), range 0.4-20.8 x 10(6)). Glucose, lactate, LDH and pH were not statistically different in the three protocols. Morphology score was best in protocol III.

CONCLUSIONS

PLT collection with the MCS 3p blood cell separator results in sufficient thrombocyte yields. Using the haemocalculator we were not able to achieve the desired platelet yield. For this reason, and because of the higher WBC contamination in protocol II we prefer PLT collection without the haemocalculator. The quality of the platelet concentrates was good with respect to the parameters glucose, lactate, LDH, morphology score and pH.