Suppr超能文献

永久性心脏起搏系统患者的射频导管房室结消融术

Radiofrequency catheter atrioventricular node ablation in patients with permanent cardiac pacing systems.

作者信息

Chang A C, McAreavey D, Tripodi D, Fananapazir L

机构信息

Cardiology Branch, National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, Maryland 20892.

出版信息

Pacing Clin Electrophysiol. 1994 Jan;17(1):65-9. doi: 10.1111/j.1540-8159.1994.tb01352.x.

Abstract

Following successful RF ablation of the atrioventricular node (AVN), temporary pacing is necessary prior to insertion of a permanent pacemaker. The risks and inconvenience of temporary pacing could be avoided if a permanent pacemaker is already in place. This study reports the feasibility of RF ablation of the AVN in 27 patients (age 55 +/- 17 years, 15 males) with hypertrophic cardiomyopathy and pacemakers. Indications for AVN ablation were drug refractory atrial fibrillation in 24 patients, and rapid AVN conduction preventing septal pre-excitation by DDD pacemaker, inserted for relief of left ventricular outflow obstruction, in three cases. Sixteen patients had DDD devices and 11 patients had VVI devices. During RF ablation, each pacemaker was programmed to VVI at 50 beats/min. The ablation catheter was manipulated with fluoroscopic control to avoid close contact with or disturbance of the pacing leads. In 16 patients, RF ablation was performed immediately following pacemaker implantation but in the remaining patients, the AVN was ablated 6-32 months after pacemaker implantation. The power applied was 25-50 watts for a duration of 15-60 seconds. AV block was achieved in all cases but required 34 +/- 36 applications for 16.5 +/- 17.8 min/case. RF ablation consistently caused reversion to magnet rate in one patient and temporarily inhibited appropriate pacemaker discharge in another. However, no other pacemaker or lead malfunction was detected so that temporary pacing was not required in any case. At 6 +/- 3 months follow-up, all pacemakers were functioning normally without alteration in pacing parameters from baseline.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

在成功进行房室结(AVN)射频消融术后,植入永久性起搏器之前需要进行临时起搏。如果已经植入了永久性起搏器,就可以避免临时起搏的风险和不便。本研究报告了对27例(年龄55±17岁,男性15例)患有肥厚型心肌病且已植入起搏器的患者进行AVN射频消融的可行性。AVN消融的指征为24例药物难治性心房颤动,以及3例因快速AVN传导妨碍了为缓解左心室流出道梗阻而植入的DDD起搏器对间隔预激的治疗。16例患者植入了DDD装置,11例患者植入了VVI装置。在射频消融期间,每个起搏器均程控为VVI模式,频率为50次/分钟。在荧光透视控制下操作消融导管,以避免与起搏导线密切接触或造成干扰。16例患者在起搏器植入后立即进行射频消融,但其余患者在起搏器植入后6 - 32个月进行AVN消融。施加的功率为25 - 50瓦,持续时间为15 - 60秒。所有病例均实现了房室传导阻滞,但平均每例需要34±36次放电,持续16.5±17.8分钟。射频消融在1例患者中持续导致起搏器转为磁频率,在另1例患者中暂时抑制了起搏器的正常放电。然而,未检测到其他起搏器或导线故障,因此在任何情况下均无需临时起搏。在6±3个月的随访中,所有起搏器功能正常,起搏参数与基线相比无变化。(摘要截短于250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验