A rapid, accurate, and sensitive method was developed for the determination of molybdenum in seawater and bovine liver samples by reversed-phase liquid chromatography (LC) using a C18 column. The method was based on chelation of the metal with 8-hydroxyquinoline by on-column or precolumn derivatization followed by LC elution with acetonitrile-0.02M acetate buffer (3 + 2; pH 4.1) and 1 x 10(-3)M of the ligand as the mobile phase. The spectrophotometric detection was made at 390 nm. Data obtained from the 2 derivatizations (on- and precolumn chelation) were compared in terms of detection limit, precision, accuracy, and ease of LC determination of molybdenum. No significant difference exists in the detection limit and the sensitivity of the 2 methods. However, the on-column chelation method, that is, the direct injection of aqueous samples onto the column, showed relatively less interference from manganese and greater simplicity in operation compared with precolumn chelation; therefore, on-column chelation is preferable for routine analyses. The on-column method produced mean levels of 9.6 +/- 0.4 micrograms/L for the seawater sample and 3.6 +/- 0.2 microgram/g (certified value, 3.5 +/- 0.5 microgram/g) for bovine liver.
建立了一种快速、准确且灵敏的方法,用于通过反相液相色谱(LC)使用C18柱测定海水和牛肝样品中的钼。该方法基于金属与8-羟基喹啉通过柱上或柱前衍生化进行螯合,然后以乙腈-0.02M醋酸盐缓冲液(3 + 2;pH 4.1)和1×10(-3)M的配体作为流动相进行LC洗脱。在390nm处进行分光光度检测。从两种衍生化方法(柱上和柱前螯合)获得的数据在检测限、精密度、准确度以及钼的LC测定简便性方面进行了比较。两种方法在检测限和灵敏度方面不存在显著差异。然而,柱上螯合法,即将水样直接进样到柱上,与柱前螯合相比,显示出受锰的干扰相对较小且操作更简便;因此,柱上螯合更适合常规分析。柱上法对海水样品测得的平均含量为9.6±0.4微克/升,对牛肝测得的平均含量为3.6±0.2微克/克(认证值为3.5±0.5微克/克)。