Spot size reduction in coupling of laser to micromanipulator in laser microsurgery by fiberoptic link.

作者信息

Logunov S L, Lemberg V G

机构信息

Medical Division, Reliant Technologies Inc., Foster City, California 94404, USA.

出版信息

Lasers Surg Med. 1995;17(1):82-5. doi: 10.1002/lsm.1900170110.

Abstract

BACKGROUND AND OBJECTIVE

In coupling laser with micromanipulator through fiberoptics, the resulting diameter of the spot is limited by the laws of geometrical optics, because of the high numerical aperture (N.A.) of fiberoptic radiation. A new method for the reduction of spot size diameter is suggested.

STUDY DESIGN/MATERIALS AND METHODS: The output of a 2 mW He-Ne laser was couped via fiberoptic link, the fiberoptic output light collected by a single lens collimator and directed to the input of the micromanipulator.

RESULTS

The spot size can be considerably reduced by the introduction of an aperture which reduces the numerical aperture (N.A.) of the fiberoptic. The resulting reduction in total power has little effect on the power density.

CONCLUSION

This approach to the collimator design permits reduction in spot size without any significant changes in power density, thus avoiding damage to the tissue and obtaining optimum performance from the micromanipulator.

摘要

相似文献

2
The effect of optical design on micromanipulator spot size using CO2 laser irradiation.
Otolaryngol Head Neck Surg. 2002 Jun;126(6):593-7. doi: 10.1067/mhn.2002.125298.
3
New microspot micromanipulator for carbon dioxide laser surgery in otolaryngology. Early clinical results.
Arch Otolaryngol Head Neck Surg. 1988 Sep;114(9):1012-5. doi: 10.1001/archotol.1988.01860210078020.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验