Friberg B, Sennerby L, Roos J, Johansson P, Strid C G, Lekholm U
Department of Handicap Research, Faculty of Medicine, University of Göteborg, Sweden.
Clin Oral Implants Res. 1995 Sep;6(3):164-71. doi: 10.1034/j.1600-0501.1995.060305.x.
A method using cutting resistance measurements during low-speed threading for identification of various bone densities has been evaluated with regard to its precision and potential. Pig ribs were used as test samples. Differing hand pressure, minor deviation (5 degrees) from a vertical tapping direction and individual threading did not reveal any significant differences in cutting resistance values. After implants were inserted into the threaded canals, the total bone as well as trabecular and compact bone areas surrounding the implants were calculated via a computer program and using microradiographs of the bone test samples. The outcome of the cutting resistance measurements was compared with that of the microradiographic technique, and good agreement was observed between the two procedures in the ability to identify bone density. Therefore, cutting resistance measurements may in the future also be used to clinically identify bone qualities in jaws.