Continuous-flow fluoroimmunoassay of serum gentamicin, with automatic sample blank correction.

作者信息

Shaw E J, Watson R A, Smith D S

出版信息

Clin Chem. 1979 Feb;25(2):322-4.

PMID:759029
Abstract

We describe the continuous-flow automation of a quenching fluoroimmunoassay for determination of the antibiotic gentamicin in serum. In the flow system, sample is mixed with fluorescein-labeled gentamicin, followed by antiserum, and after a short incubation the fluorescence is measured. The intrinsic blank signal of each sample is corrected for during its single passage through the system by pumping antiserum discontinuously, so as to divide the final stream into sections corresponding to conventional separate assay and blank mixtures. Analyses time for each sample is 6 min, and 20 samples can be run per hour. Sensitivity, precision, and accuracy are adequate for clinical purposes, and assays of patients' samples correlate satisfactorily with the results of manual quenching and polarization fluoroimmunoassays, and plate diffusion bioassay.

摘要

我们描述了一种用于测定血清中抗生素庆大霉素的淬灭荧光免疫分析的连续流动自动化方法。在流动系统中,样品与荧光素标记的庆大霉素混合,随后加入抗血清,短暂孵育后测量荧光。每个样品的固有空白信号在其单次通过系统时通过不连续泵送抗血清进行校正,以便将最终液流分成对应于传统单独测定和空白混合物的部分。每个样品的分析时间为6分钟,每小时可进行20个样品的检测。灵敏度、精密度和准确度足以满足临床目的,患者样品的检测结果与手动淬灭和偏振荧光免疫分析以及平板扩散生物测定的结果具有良好的相关性。