Watson R A, Landon J, Shaw E J, Smith D S
Clin Chim Acta. 1976 Nov 15;73(1):51-5. doi: 10.1016/0009-8981(76)90303-x.
A polarisation fluoroimmunoassay has been developed for the routine determination of gentamicin levels in serum. The method employs fluorescein-labelled gentamicin which is easily, reproducibly and economically prepared and has excellent shelf-life. The assay is fast (minimum incubation time 2 min) and requires 1.25 mul of serum. There is no separation step prior to readout of results. Recovery experiments reveal no serum effects. Intra- and inter-assay coefficients of variation are of the order of 10% or less. Estimations of gentamicin levels in patient serum samples correlate closely with those obtained by bioassay (r = 0.93) and radioimmunoassy (r = 0.97).