Dynamics of a mechanical monoleaflet heart valve prosthesis in the closing phase: effect of squeeze film.

作者信息

Gill-Jeong C, Chandran K B

机构信息

Department of Mechanical Engineering, Won-Kwang University, Iri city, Jeon-Buk, Korea.

出版信息

Ann Biomed Eng. 1995 Mar-Apr;23(2):189-97. doi: 10.1007/BF02368325.

Abstract

An analysis of the dynamics of a mechanical monoleaflet heart valve prosthesis in the closing phase is presented. The backflow velocity of the fluid and the pressure distribution on the occluder during the closing phase were computed using a control volume approach in the unsteady state. Using moment equilibrium principles on the occluder motion and the squeeze film dynamics of the fluid between the occluder and the guiding strut at the instant of impact, the velocity of the occluder tip and the impact force between the occluder and the guiding struts were computed. The dynamics of fluid being squeezed between the occluder and the guiding struts are accounted for by Reynolds' equation. The effect of the fluid being squeezed between the occluder and the guiding strut was to reduce the velocity of the occluder tip at the instant of valve closure as well as to dampen the fluttering of the occluder before coming to rest in the fully closed position. The squeeze film fluid pressure changed rapidly from a high positive value (10 MPa) to a relatively large negative value (-15 MPa) in < 1 msec. The results of this study may be extended for the analysis of cavitation inception and mechanical stresses on the formed elements and valve components, as well as to estimate the endurance limits of prosthetic valves.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验