Suppr超能文献

评估由另一种聚酯纱线来源编织而成的Dialine血管假体。

Evaluating the Dialine vascular prosthesis knitted from an alternative source of polyester yarns.

作者信息

King M W, Marois Y, Guidoin R, Ukpabi P, Deng X, Martin L, Pâris E, Douville Y

机构信息

Department of Surgery, Laval University, Quebec City, Canada.

出版信息

J Biomed Mater Res. 1995 May;29(5):595-610. doi: 10.1002/jbm.820290507.

Abstract

The sudden and unilateral decision by E. I. DuPont de Nemours & Co., Inc., to withdraw its polymers for use in implantable devices has presented the medical device industry with an immediate and serious challenge to find alternative sources of biomaterials. In France, the company Cardial S.A. has already taken steps to find an alternative polyester yarn to replace Dacron by developing a new arterial prosthesis knitted from polyester yarns supplied by Rhône-Poulenc Fibres. This article describes an in vitro and in vivo study of this French device, called the Dialine prosthesis, with a view to determining its relative performance compared to current American and British prostheses, which rely on DuPont's Dacron yarn. In addition to analyses of the morphology and textile structure, and measurements of its physical and chemical properties, the Dialine graft was implanted as a thoracoabdominal bypass in dogs for periods ranging from 4 h to 6 months. In addition to our pathologic and histologic observations, we cleaned and evaluated the explanted prostheses for in vivo changes in dimensions, strength, and crystalline microstructure. The Dialine graft was found to differ structurally from other polyester prostheses because it is warp-knitted from a mixture of flat and texturized yarns with finer filaments. Its denser structure has a lower water permeability, greater flexibility and ease of handling, satisfactory strength, and dimensional stability, and it presents different textures on its luminal and external surfaces. The in vivo trial demonstrated that it has excellent biocompatibility and biostability over 6 months. With no thrombi observed on the luminal surface after 3 months, it has a faster rate of healing, generates compact external and internal capsules with a thinner neointima, and has an overall milder inflammatory response than is normally observed with Dacron-based prostheses.

摘要

伊士曼·柯达公司单方面突然决定停止供应用于植入式设备的聚合物,这给医疗器械行业带来了一项紧迫而严峻的挑战,即寻找生物材料的替代来源。在法国,卡迪亚尔公司已采取措施,通过开发一种由罗纳-普朗克纤维公司提供的聚酯纱线编织而成的新型动脉假体,来寻找替代涤纶的聚酯纱线。本文描述了对这种名为Dialine假体的法国设备进行的体外和体内研究,旨在确定其与目前依赖杜邦涤纶纱线的美国和英国假体相比的相对性能。除了对其形态和纺织结构进行分析以及对其物理和化学性质进行测量外,还将Dialine移植物作为胸腹旁路移植到狗体内,时间从4小时到6个月不等。除了我们的病理和组织学观察外,我们还对取出的假体进行了清洁,并评估了其在体内尺寸、强度和晶体微观结构的变化。结果发现,Dialine移植物在结构上与其他聚酯假体不同,因为它是由扁平纱线和变形纱线混合而成的较细纤维长丝经编而成。其更致密的结构具有更低的透水性、更大的柔韧性和易操作性、令人满意的强度和尺寸稳定性,并且其管腔和外表面呈现出不同的纹理。体内试验表明,它在6个月内具有出色的生物相容性和生物稳定性。3个月后,在管腔表面未观察到血栓,其愈合速度更快,形成的外部和内部包膜紧密,内膜较薄,并且与基于涤纶的假体相比,炎症反应总体上更轻微。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验