[实验性超声血管成形术:血栓的体外溶解]

[Experimental ultrasound angioplasty: in vitro resolution of thrombi].

作者信息

Stähr P, Erbel R, Weber W, Fischer H, Meyer J

机构信息

II. Medizinische Klinik und Poliklinik Johannes-Gutenberg-Universität, Mainz.

出版信息

Z Kardiol. 1995 May;84(5):365-72.

PMID:7625098
Abstract

A new ultrasonic angioplasty ablation catheter connected to a 19.5 kHz. 25 W transducer was tested in vitro for its ability to disrupt 12-h. 24-h, and 5-day-old whole blood thrombi (n = 45.697 mg +/- 223 mg) and fibrin thrombi (n = 45.338 mg +/- 133 mg), as well as 5-day-old cadaver thrombi (n = 8.270 mg +/- 71 mg) within 10 min. Five of each age were used as control thrombi in which the catheter was moved back and forth without ultrasound emission. The size of ablated thrombus particles was measured by a laser device. The power output at the end of the catheter was assessed calorimetrically. The loss of weight of whole blood thrombi was between 429 (74%) and 524 mg (91%) (p < 0.01, whole blood thrombi vs. control thrombi) and between 302 (85%) and 314 mg (95%) (p < 0.05) for fibrin thrombi, respectively. Thrombus age did not prove to be a highly significant influencing factor. The disruption rate for whole blood thrombi was 0.75 to 1.05 mg/s and for fibrin thrombi 0.69 to 0.7 mg/s. It was only 0.09 mg/s for the cadaver thrombi. 93% of all particles ablated from whole blood thrombus ranged between 0-5 micron, less than 0.2% between 30-150 microns. For fibrin thrombi 69% of all particles were < 10 microns (25% between 10-20 microns). Only 0.02% ranged between 300-600 microns, which was similar for cadaver thrombi. The mean measured power output at the catheter tip was 5.9 W compared to the power output of 25 W at the ultrasound generator.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

一种连接到19.5千赫、25瓦换能器的新型超声血管成形术消融导管,在体外测试其破坏12小时、24小时和5日龄全血血栓(n = 45,697毫克±223毫克)和纤维蛋白血栓(n = 45,338毫克±133毫克)以及5日龄尸体血栓(n = 8,270毫克±71毫克)的能力,时间为10分钟。每种年龄的血栓各取5个作为对照血栓,导管在不发射超声的情况下前后移动。通过激光装置测量消融血栓颗粒的大小。用热量测定法评估导管末端的功率输出。全血血栓的重量减轻分别在429毫克(74%)至524毫克(91%)之间(p < 0.01,全血血栓与对照血栓相比),纤维蛋白血栓在302毫克(85%)至314毫克(95%)之间(p < 0.05)。血栓年龄并非高度显著的影响因素。全血血栓的破坏速率为0.75至1.05毫克/秒,纤维蛋白血栓为0.69至0.7毫克/秒。尸体血栓仅为0.09毫克/秒。从全血血栓中消融的所有颗粒中,93%的范围在0 - 5微米之间,30 - 150微米之间的不到0.2%。对于纤维蛋白血栓,所有颗粒的69% < 10微米(25%在10 - 20微米之间)。只有0.02%在300 - 600微米之间,尸体血栓情况类似。导管尖端测得的平均功率输出为5.9瓦,而超声发生器的功率输出为25瓦。(摘要截选至250字)

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验