液体潴留对使用生物电阻抗法测量身体成分的影响。

Effects of fluid retention on the measurement of body composition using bioelectric impedance.

作者信息

Wu T J, Huang J J, Lin C Y

机构信息

Department of Internal Medicine, National Cheng Kung University Hospital, Tianan, Taiwan, R.O.C.

出版信息

J Formos Med Assoc. 1994 Nov-Dec;93(11-12):939-43.

PMID:7633198
Abstract

Body composition measurements using bioelectrical impedance analysis (BIA) were performed on 50 uremic patients immediately before and after hemodialysis (HD) therapy, on 10 uremic patients under continuous ambulatory peritoneal dialysis (CAPD) therapy before and after the fluid was drained out of the peritoneal cavity and on 3 cirrhotic patients before and after abdominal paracentesis. Thirty-two pairs of measurements were performed on the CAPD patients and 3 pairs on the cirrhotic patients. Significant increases in both resistance and reactance were noted after HD. However, resistance and reactance before and after peritoneal fluid had been drained were not significantly different. In HD, body fluid loss as estimated by BIA (LBIA) was higher than that from the body weight method (LBW). The relationship between the two is described by the equation LBW = 0.841LBIA + 0.588. The total body fat estimated before HD was significantly lower than after HD, and it was significantly higher when measured with peritoneal fluid retention than with fluid drained. Generalized edema caused an increase in resistance and reactance, while local peritoneal fluid retention did not cause any significant changes. When total body fluid is measured by the BIA method, it is often overestimated in the case of generalized edema and underestimated in case of peritoneal fluid retention. Therefore, total body fat is underestimated by BIA in generalized edema before HD and overestimated in peritoneal fluid retention.

摘要

采用生物电阻抗分析(BIA)对50例尿毒症患者在血液透析(HD)治疗前后、10例持续性非卧床腹膜透析(CAPD)治疗患者在腹腔引流前后以及3例肝硬化患者在腹腔穿刺前后进行身体成分测量。对CAPD患者进行了32对测量,对肝硬化患者进行了3对测量。HD后电阻和电抗均显著增加。然而,腹腔引流前后的电阻和电抗无显著差异。在HD中,通过BIA估计的体液丢失(LBIA)高于体重法估计的体液丢失(LBW)。两者之间的关系由方程LBW = 0.841LBIA + 0.588描述。HD前估计的全身脂肪显著低于HD后,且腹腔积液时测量的全身脂肪显著高于引流后。全身性水肿导致电阻和电抗增加,而局部腹腔积液未引起任何显著变化。当通过BIA方法测量全身液体时,在全身性水肿情况下常被高估,在腹腔积液情况下常被低估。因此,HD前全身性水肿时BIA低估了全身脂肪,而在腹腔积液时高估了全身脂肪。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验