Suppr超能文献

Automated and accurate assessment of the distribution, magnitude, and direction of pincushion distortion in angiographic images.

作者信息

van der Zwet P M, Meyer D J, Reiber J H

机构信息

Department of Diagnostic Radiology and Nuclear Medicine, University Hospital Leiden, The Netherlands.

出版信息

Invest Radiol. 1995 Apr;30(4):204-13. doi: 10.1097/00004424-199504000-00002.

Abstract

RATIONALE AND OBJECTIVES

Pincushion distortion continues to be a potential problem for the accurate assessment of arterial and catheter dimensions from x-ray angiograms. The authors investigate whether the distortion of state-of-the-art intensifiers is yet small enough to be neglected, and whether the rotation/angulation of the x-ray system plays a significant role.

METHODS

The location and degree of distortion from x-ray images of a centimeter grid, which is positioned against the input screen of the image intensifier, are assessed automatically using image processing techniques. A value for the maximum amount of change in the distortion vector field is derived that allows the estimation of the maximum relative error associated with a diameter measurement uncorrected for pincushion distortion.

RESULTS

The accuracy of the algorithm itself was assessed by rotating and translating the centimeter grid under the image intensifier at anteroposterior position. For the distortion vector length, the standard deviation in the measurement of the distortion areas was found to be 3.7 cm2 (1.3% of the total area). For the gradient values, the standard deviation was 2.2 cm2 or 0.75% of the total image intensifier area. In the second evaluation study, the centimeter grid was fixed onto the input screen of the image intensifier, and the gantry was rotated to span all possible positions of the system. In this case, the changes in measured areas were often much larger (up to 51.25 cm2 for a 9-inch image intensifier, equivalent to 15.6% of the total image intensifier area) than the standard deviations that had been found in the first evaluation study.

CONCLUSIONS

The distortion is highly dependent upon the actual spatial position of the image intensifier, and correcting for pincushion distortion may therefore introduce larger errors than leaving the measurements uncorrected.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验