Gowrie-Mohan S, Chakrabarti M K, Lockwood G G
Department of Anaesthesia, Hammersmith Hospital, London.
Anaesthesia. 1995 Aug;50(8):706-8. doi: 10.1111/j.1365-2044.1995.tb06099.x.
We have examined the predictability of inspired isoflurane concentration during low-flow anaesthesia using a to-and-fro breathing system. Twenty one adult patients requiring mechanical ventilation of the lungs during surgery took part in this study. Using a fresh gas flow of 2 l.min-1, the ratio of inspired isoflurane concentration to isoflurane vaporizer setting was found to be approximately 4/5th after 5 min of anaesthesia. The ratio was maintained throughout the procedure, except for a few minutes following each change in vaporizer setting.