Suppr超能文献

Doppler guided extrathoracic introducer insertion.

作者信息

Fyke F E

机构信息

Mississippi Baptist Medical Center, Jackson, USA.

出版信息

Pacing Clin Electrophysiol. 1995 May;18(5 Pt 1):1017-21. doi: 10.1111/j.1540-8159.1995.tb04742.x.

Abstract

Recent reports have raised concern over the use of standard subclavian puncture for placement of permanent pacing leads. This study reports the initial experience of one implanter in 59 consecutive, unselected patients undergoing implantation of a variety of pacing leads in whom a simple Doppler flow detector was used to guide extrathoracic venipuncture of the lateral subclavian/axillary vein. A total of 100 leads were placed for 48 right- and 11 left-sided systems. Lead insertion by this technique was successful in all patients. One small pneumothorax and two uncomplicated arterial punctures occurred. One patient's leads were revised electively because of unacceptable loss of redundancy when standing. There have been no infections or other late complications. A simple Doppler flow detector appears to be useful in identifying the lateral subclavian/axillary vein for extrathoracic pacemaker lead placement.