Am J Clin Pathol. 1976 Feb;65(2):242-9. doi: 10.1093/ajcp/65.2.242.
The presence of small gas bubbles in devices used to collect blood gas samples is considered. A gas bubble whose relative volume is 0.5 to 1% or more that of the liquid in the collection device is a potential source of significant error. Relationships describing the possible magnitude of this error are derived and substantiated.