Kawchuk G, Herzog W
Human Performance Laboratory, University of Calgary, Alberta, Canada.
J Manipulative Physiol Ther. 1995 Jun;18(5):298-301.
To assess the reliability and the accuracy of a common method of tissue compliance measurement using a series of non-biological test surfaces (foam). Currently, tissue compliance measurement is most commonly obtained with a hand-held instrument known as a tissue compliance meter (TCM).
Descriptive study.
Human Performance Laboratory, University of Calgary.
A TCM was tested on four surfaces (three test and one control) with five different input forces resulting in 20 unique surface/force combinations. For each combination of surface/force, ten trials were obtained in a random order by each of five examiners, yielding a total of 1000 separate measurements of surface compliance.
Millimeters of surface displacement per Newtons of input force.
Intraclass correlation coefficients were calculated for each of the 20 surface/force combinations to judge interexaminer reliability. The median of these 20 coefficients was 0.005: the greatest single value tending toward complete reliability was 0.22. Trials obtained from the control surface (assumed to be incompressible), demonstrated a range of displacements from 0.00 to 2.00 mm.
Within the design of this experiment, the reliability and accuracy of the hand-held tissue compliance meter was poor. We would suggest that the adequacy of this instrument in clinical practice or scientific work must be seriously questioned. Although the assessment of tissue compliance may be useful in the characterization of the musculoskeletal system and particularly in the assessment of treatment outcome, we surmise that a more reliable and accurate instrument is needed for the quantification of tissue compliance.
使用一系列非生物测试表面(泡沫)评估一种常用组织顺应性测量方法的可靠性和准确性。目前,组织顺应性测量最常用一种称为组织顺应性测量仪(TCM)的手持仪器来进行。
描述性研究。
卡尔加里大学人体性能实验室。
使用TCM在四个表面(三个测试表面和一个对照表面)上施加五种不同的输入力,从而产生20种独特的表面/力组合。对于每种表面/力组合,由五名检查人员以随机顺序进行十次试验,总共获得1000次表面顺应性的单独测量值。
每牛顿输入力作用下表面位移的毫米数。
计算20种表面/力组合中每种组合的组内相关系数,以判断检查人员之间的可靠性。这20个系数的中位数为0.005:最接近完全可靠性的单个值为0.22。从对照表面(假定为不可压缩)获得的试验显示位移范围为0.00至2.00毫米。
在本实验设计范围内,手持组织顺应性测量仪的可靠性和准确性较差。我们认为,该仪器在临床实践或科研工作中的适用性必须受到严重质疑。尽管组织顺应性评估在肌肉骨骼系统特征描述中可能有用,特别是在治疗效果评估中,但我们推测需要一种更可靠、准确的仪器来量化组织顺应性。