Breum J, Wiberg J, Bolton J E
Anglo-European College of Chiropractice, Bournemouth, England.
J Manipulative Physiol Ther. 1995 Oct;18(8):497-502.
To establish the reliability of the BROM II device for measuring lumbar mobility in the sagittal, coronal and transverse planes and its validity against the double inclinometer method.
Blind intra- and interexaminer reliability and concurrent validity. Interexaminer reliability was determined between two examiners.
Chiropractice teaching college.
Forty-seven asymptomatic chiropractic students (27 men and 20 women, age range 18 to 38 yr).
Lumbar mobility measurement in degrees using the BROM II and double inclinometer techniques.
Intraclass correlation coefficients (ICC) showed good intra- and interexaminer reliability of the BROM II for flexion (0.91 and 0.77, respectively) and lateral flexion (0.91 and 0.85 respectively). Less support was given to the reliability of the instrument in extension (0.63 and 0.35, respectively) and rotation (0.57 and 0.36 respectively). Concurrent validity of the BROM II and double inclinometer methods was partially supported (ICC in all planes range from 0.27 to 0.75).
The BROM II was found to be a reliable instrument in the measurement of lumbar mobility in the sagittal (flexion) and coronal planes. However, before this device can be recommended as an assessment tool in clinical practice or clinical trials, further investigation into its reliability in a symptomatic group of patients is required. (J Manipulative Physiol Ther 1995; 18:497-502).
确定BROM II设备测量腰椎矢状面、冠状面和横断面活动度的可靠性及其与双倾角计法相比的有效性。
检查者内和检查者间的盲法可靠性及同时效度。检查者间可靠性由两名检查者确定。
整脊疗法教学学院。
47名无症状的整脊疗法专业学生(27名男性和20名女性,年龄范围18至38岁)。
使用BROM II和双倾角计技术以度为单位测量腰椎活动度。
组内相关系数(ICC)显示,BROM II在测量前屈(分别为0.91和0.77)和侧屈(分别为0.91和0.85)时,检查者内和检查者间具有良好的可靠性。对于该仪器在伸展(分别为0.63和0.35)和旋转(分别为0.57和0.36)时的可靠性支持较少。BROM II和双倾角计法的同时效度得到部分支持(所有平面的ICC范围为0.27至0.75)。
发现BROM II是测量腰椎矢状面(前屈)和冠状面活动度的可靠仪器。然而,在该设备被推荐用于临床实践或临床试验的评估工具之前,需要对有症状患者群体中其可靠性进行进一步研究。(《手法治疗与生理疗法杂志》1995年;18:497 - 502)