Electromagnetic signals coming from the pacemaker programmer can permanently damage multiprogrammable pacemakers. At a follow-up of a sixfold programmable VVI pacemaker (Phoenix 251-6, Pacesetter), the emergency VVI button was activated on the programmer after the programming of the pacing impulse had induced a loss of capture. The next telemetrical interrogation revealed an alteration to a threefold programmable Programmalith III. The inadvertent programming of the model number occurred, because the programming head was removed at one very specific time in the programming sequence of the emergency VVI command.
来自起搏器程控仪的电磁信号可能会永久性损坏多程控功能起搏器。在对一款六程控功能VVI起搏器(Pacesetter公司的Phoenix 251 - 6)进行随访时,在设置起搏脉冲导致夺获失败后,在程控仪上激活了紧急VVI按钮。下一次遥测问询显示该起搏器已变为三程控功能的Programmalith III型。出现型号误程控的情况,是因为在紧急VVI命令的程控序列中一个非常特定的时间移除了程控头。