Del Bufalo A G, Schlaepfer J, Fromer M, Kappenberger L
Division of Cardiology, CHUV, Lausanne, Switzerland.
Pacing Clin Electrophysiol. 1993 Jun;16(6):1240-4. doi: 10.1111/j.1540-8159.1993.tb01709.x.
Efforts have been made to design electrodes that significantly reduce not only the acute and chronic stimulation thresholds, but also attenuate the early peaking phenomenon and polarization. At two voltage levels (2.7 V and 5.4 V, respectively), we evaluated the right ventricular stimulation thresholds obtained with a new, iridium oxide-coated electrode in ten patients who received a VVI pacemaker. Measurements were made at implant and at multiple intervals for 1 year. Pulse width stimulation thresholds at implant were as follow: 0.04 +/- 0.008 msec at 2.7 V, 0.03 +/- 0.004 msec at 5.4 V; values at 2 weeks were 0.14 +/- 0.06 msec at 2.7 V, 0.07 +/- 0.025 msec at 5.4 V; values at 3 months were 0.09 +/- 0.03 msec at 2.7 V, 0.05 +/- 0.01 msec at 5.4 V; values at 1 year were 0.08 +/- 0.02 msec at 2.7 V, 0.04 +/- 0.01 msec at 5.4 V. The maximal increase of 0.11 +/- 0.05 msec occurred at 2.7 V, 2 weeks after implant. Our results indicate that this new electrode provides low acute and long-term stimulation thresholds, as well as an attenuated early peaking phenomenon, being able to stimulate safely at 2.7 V even early after implant.
人们已努力设计出不仅能显著降低急性和慢性刺激阈值,还能减弱早期峰值现象和极化的电极。在两个电压水平(分别为2.7 V和5.4 V)下,我们评估了10名接受VVI起搏器治疗的患者使用新型氧化铱涂层电极获得的右心室刺激阈值。在植入时以及术后1年的多个时间间隔进行测量。植入时的脉宽刺激阈值如下:2.7 V时为0.04±0.008毫秒,5.4 V时为0.03±0.004毫秒;2周时的值为2.7 V时0.14±0.06毫秒,5.4 V时0.07±0.025毫秒;3个月时的值为2.7 V时0.09±0.03毫秒,5.4 V时0.05±0.01毫秒;1年时的值为2.7 V时0.08±0.02毫秒,5.4 V时0.04±0.01毫秒。植入后2周,2.7 V时最大增加0.11±0.05毫秒。我们的结果表明,这种新型电极提供了较低的急性和长期刺激阈值,以及减弱的早期峰值现象,即使在植入后早期也能在2.7 V安全刺激。