Gabriel S, Atterhög J H, Orö L, Ekelund L G
Scand J Clin Lab Invest. 1976 Jan;36(1):29-34. doi: 10.1080/00365517609068015.
The cardiac output was measured simultaneously by the impedance cardiography and dye dilution methods in 10 patients with acute myocardial infarction 2-3 weeks after admission to the Coronary Care Unit. The impedance cardiac output was on the average 9.7% higher than the dye dilution cardiac output. The reproducibility of impedance cardiac output 4.1%, compared with 5.1% for the dye dilution method. The study showed a satisfactory reliability of impedance in predicting the relative changed of cardiac output in response to tilting from the supine to the 30 degrees head-up position, to a 10 degrees head-down position, and to the intravenous administration of propranolol.