Suppappola S, Sun Y
University of Rhode Island, Department of Electrical Engineering, Kingston 02881-0805, USA.
Biomed Instrum Technol. 1995 Jan-Feb;29(1):41-9.
This paper presents a QRS-detector error logger and the implementation of a fully automated system for evaluation of QRS-detector performance. The error logger requires a known reference, such as an electrocardiogram (ECG) database, that provides ECG data and reference pulses wherever QRS complexes occur. These reference pulses are used to validate detection pulses generated by a QRS detector in real time. Using an eight-bit microcontroller chip, the error logger was constructed at a very low materials cost. Applications include automated analysis of QRS-detector performance and optimization of parameters in QRS-detection algorithms.
本文介绍了一种QRS波检测器错误记录器以及用于评估QRS波检测器性能的全自动系统的实现。该错误记录器需要一个已知的参考,例如心电图(ECG)数据库,该数据库在QRS波群出现的任何地方提供ECG数据和参考脉冲。这些参考脉冲用于实时验证QRS波检测器生成的检测脉冲。使用八位微控制器芯片,以非常低的材料成本构建了错误记录器。其应用包括QRS波检测器性能的自动分析以及QRS检测算法中参数的优化。