Iwasaki A
School of Allied Medical Sciences, Hirosaki University, Aomori, Japan.
Phys Med Biol. 1995 Mar;40(3):383-91. doi: 10.1088/0031-9155/40/3/004.
A method is reported that evaluates the wall correction factor (Awall)water of an ionization chamber using measured data. The method needs a revised type of tissue-air ratio (TARcav) that should be measured with the ionization chamber. An experimental study has been carried out for 60Co gamma-rays using zero-area TARcav data extrapolated from non-zero-area TARcav data using the extrapolation method developed by Bjärngard and Petti. Using this technique, it is found that the value of (Awall)water for an ionization chamber has a deviation of -0.1% to -0.4% as compared with the Monte Carlo results. An experimental study has also been carried out using a published set of 60Co traditional tissue-air ratio (TAR) data assuming TARcav = TAR.