Goodie D B, Philip J H
Department of Anesthesia, Harvard Medical School, Brigham and Women's Hospital, Boston, MA 02115, USA.
J Clin Monit. 1995 Jan;11(1):47-50. doi: 10.1007/BF01627420.
The objective of our study was to determine if clinical observation of pressure-flow relationships (PFR) can differentiate between partial external obstruction (obstruction) and infiltration as a cause of poor performance of gravity-fed infusions.
A total of 24 patients with functional intravenous cannulae in situ had obstruction simulated by the application of a tourniquet proximal to the cannula. The change in flow (delta F) for a discrete change in pressure (delta P) was determined in each case by counting drop rates at two different elevations of the fluid reservoir level, 10 cm apart. The same process was repeated in 15 patients in whom the cannula was in an extra vascular location (infiltration). Three sizes of cannula--16-gauge, 18-gauge, and 20-gauge--were examined, with equal distribution of sizes in each group. The effect on flow rates of inflating a blood pressure (BP) cuff proximally on the cannulated limb was assessed. The ratio delta P/delta F is the total resistance of the infusion system, and by subtracting known values for resistance of infusion tubing and cannula, the venous or tissue resistance was calculated.
There was a statistically significant difference between the change in flow for obstructed compared with infiltrated cannulae for the same change in pressure for each cannula size. The mean venous resistance was 23 mm Hg/L/hr, while that of tissue was 280 mm Hg/L/hr, with no overlap between groups. There was no effect on flow rate with blood pressure cuff inflation in the infiltrated group whereas flow progressively fell in the obstructed group.
Clinical observation of PFRs in poorly functioning gravity-fed IV infusions can assist in detecting infiltration as a cause. Inflation of a blood pressure cuff will further impair flow where the cannula is intravascular, but will have no effect in an extravascular location.
我们研究的目的是确定通过临床观察压力 - 流量关系(PFR)能否区分重力输液效果不佳是由部分外部阻塞(梗阻)还是浸润引起的。
共有24例留置功能性静脉套管的患者,通过在套管近端应用止血带来模拟梗阻。在每种情况下,通过计算储液罐水平相差10 cm的两个不同高度处的滴速,确定压力离散变化(ΔP)时的流量变化(ΔF)。对15例套管位于血管外位置(浸润)的患者重复相同过程。检查了三种尺寸的套管——16号、18号和20号——每组中尺寸均匀分布。评估了在置管肢体近端充气血压袖带对流速的影响。ΔP/ΔF比值是输液系统的总阻力,通过减去输液管和套管的已知阻力值来计算静脉或组织阻力。
对于每种套管尺寸,在相同压力变化下,梗阻套管与浸润套管的流量变化之间存在统计学上的显著差异。平均静脉阻力为23 mmHg/L/hr,而组织阻力为280 mmHg/L/hr,两组之间无重叠。浸润组中血压袖带充气对流速无影响,而梗阻组中流速逐渐下降。
对重力输液功能不佳时的PFR进行临床观察有助于检测出浸润这一原因。在套管位于血管内时,血压袖带充气会进一步损害流速,但在血管外位置则无影响。